Induktasi Meter
Construction...
The circuits may be constructed on a small perf board (Radio Shack
dual mini-board (#276-168), or if you prefer, on a PCB (Printed Circuit Board).
Layout is non-critical - almost any construction technique will suffice.
Wire-wrapping or point-to-point soldering may be used.
Desciption...
The schematic shown below converts an unknown inductance into a
voltage that can be displayed on a DVM or VOM. Values between 3uH and 500uH are
measured on the L (low) range and from 100uH to 7mH on the H (high) range. NAND
gate
They are inverted by ICD (pin 11) and integrated by R4-C2 to produce
a steady dc voltage at the + output terminal. The resulting dc voltage is
proportional to LX and the repetition rate of the oscillator. R6 and R7 are
used to calibrate the unit by setting a repetition rate that produces a dc
voltage corresponding to the unknown inductance. D1 provides a 0.7 volt
constant voltage source that is scaled by R1 to produce a small offset
reference voltage for zeroing the meter on the L (low) inductance range.
When SW1 is L (low), mV corresponds to uH, and when H (high), mV
corresponds to mH. A sensitive VOM may be substituted for the DVM with a
sacrifice in resolution.
Test and Calibration...
Short the LX terminals with a piece of wire and connect a DVM set to
the 200-mV range to the output. Adjust R1 for a zero reading. Remove the short
and substitute a known inductor of approximately 400uH. Set SW1 to the L (low)
position and adjust R7 for a reading equal to the unkown inductance. Switch SW1
to the H (high) position and connect a known inductor of about 5mH. Adjust R6
for the corresponding value. For instance, if the actual value of the
calibration inductor is 4.76mH, adjust R7 so the DVM reads 476mV.
Inductance Meter
Schematic
|
|
All components are 10% tolerance. 1N4146 or equivalent may be
substituted for D1. An LM7805 may be substituted for the 78L05. All fixed
resistors are 1/4 watt carbon composition. Capacitors are in uF. R3 value may
need to be increased or decreased slightly if calibration cannot be achieved as
described in text.